S E C T I O N 6
•
O p e r a t i o n
16
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
To clear this mode, check that the Twin-Tube Connector on the
HOTLINE
Fluid Warming Set is firmly inserted in the socket.
Add Recirculating Solution Mode
The Add Recirculating Solution mode indicates that the solution level
in the reservoir is below its minimum level.
Mode characteristics
• The green Operating LED on the Display Panel turns off.
• The red Add Solution LED (
a
) on the Display Panel illuminates.
• The audible alarm sounds and repeats approximately every two
seconds.
• The recirculating solution stops circulating.
To clear this mode, add recirculating solution to the reservoir.
Over Temperature Alarm Mode
The Over Temperature Alarm mode indicates that the temperature of
the recirculating solution is at or above 41.2°C.
Mode characteristics
• The green Operating LED on the Display Panel turns off.
• The red Over Temperature LED (
b
) on the Display Panel
illuminates.
• The audible alarm sounds and repeats approximately every two
seconds.
• The recirculating solution stops circulating.
For instructions to clear this mode, see Section 8,
Troubleshooting
.
a
b
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...