Lubricate O-Ring Seals
1
Place a small amount of silicone lubricant on a cotton swab.
2
Apply silicone lubricant along the O-Rings inside the socket (
a
)
located on the right side of the
HOTLINE
Warmer.
Silicone lubricant is available from Smiths Medical, (Silicone
lubricant P/N 80-04-002).
Maintenance Performed Every 12 Months
Disinfect the Reservoir and Change Recirculating Solution for
0.3% Hydrogen Peroxide Solution
Refer to
Disinfect the Reservoir and Change the Recirculating Solution
procedure in this section.
Replace O-Rings
(O-Ring Kit: P/N 80-04-001)
1
Unplug the
HOTLINE
Warmer before servicing.
2
Remove the socket head screws (
b
) with a 0.31cm (1/8") hex
wrench.
3
Remove the face plate, being careful not to damage the micro-
switch lever.
4
Remove the old O-rings and clean the sockets with a cotton swab.
5
Apply silicone lubricant to two new O-rings and install into the
sockets.
6
Reassemble in reverse order, being careful not to damage the micro-
switch lever.
7
Insert
HOTLINE
Fluid Warming Set and power on
HOTLINE
Warmer to verify that there are no leaks around the face plate.
Testing HOTLINE
Warmer Operation
Perform all the tests described in the testing section of this manual. See
Section 9,
Testing
. The Scheduled Maintenance Checklist below also lists
the tests.
o
o
C
C
HOTLINE
Fluid Warmer
a
o
o
C
C
HOTLINE
Fluid Warmer
b
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...