S E C T I O N 3
•
I m p o r t a n t S a f e t y I n f o r m a t i o n
6
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
Continued from previous page
WARNINGS
WARNINGS
• Prime the recirculating solution path before connecting to the
intravenous extension set. This is to confirm that there is not a
breach between the recirculating solution path and intravenous
path. If fluid exits the patient end of the HOTLINE
®
Fluid
Warming Set before connecting to the intravenous
administration set, remove and replace HOTLINE
®
Fluid
Warming Set.
• Remove all air from the HOTLINE
®
Fluid Warming Set, L-10, PC-
8, and YC-8 before connecting to the patient. Failure to do so
may result in introduction of air to the patient.
• To reduce the risk of outgassed microbubbles entering patient
vasculature, an L-10 Gas Vent may be used with the HOTLINE
®
Fluid Warming Set.
• Not for use with pressure devices generating over 300 mmHg.
Pressure greater than 300 mmHg may compromise the integrity
of the HOTLINE
®
Fluid Warming Set.
• To prevent a breach between the recirculating solution path and
intravenous path, do not use needles greater than 38 mm (1.5")
in length when accessing the injection port. If there is a breach
between the recirculating solution path and intravenous path,
patient illness may occur because of the HOTLINE
®
Warmer's
recirculating solution entering the patient's bloodstream.
• Do not stick the HOTLINE
®
Fluid Warming Set with needles, as
this will breach the I.V. path and compromise the integrity of the
patient intravenous line. If a Disposable Set with a breached
recirculating solution path/intravenous path is used, then
patient illness may occur because of the HOTLINE
®
Warmer's
recirculating solution entering the patient's bloodstream.
• Activation of the Over Temperature warning signal indicates that
warming has stopped and immediate operator intervention is
required to clear the over temperature condition or to remove
the device from service.
• If any visual indicator does not illuminate or the audible signal
does not sound, do not use the HOTLINE
®
Warmer. Remove the
device from service immediately.
• Do not operate the HOTLINE
®
Warmer in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
The risk of explosion exists if the HOTLINE
®
Warmer is operated
in a potentially explosive environment.
Continued
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...