S E C T I O N 9
•
T e s t i n g
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
27
Periodic Electrical Testing
Electrical Safety Tests must be performed by qualified personnel authorized
by the institution to perform such testing. The Safety Tests must be
performed and documented at least once per year, or according to
institutional policy. These tests include but are not limited to:
• Leakage current
• Ground bond test
Leakage Current
Leakage current must be tested according to methods and pass/fail
criteria described in IEC 60601-1:2005. Leakage current must be
performed with the heater circuit in the full ON condition. To achieve
this condition, perform the test when the reservoir is at room
temperature. When the HOTLINE
Warmer is first turned on and the
temperature is rapidly rising, but still below 39°C, the heater circuit is
in a full ON condition.
Note: The HOTLINE
Warmer is equipped with sensing interlocks. A
HOTLINE
Fluid Warming Set is required to correctly operate the
HOTLINE
Warmer and perform leakage current testing. Do not defeat
the sensing interlocks or try to operate the HOTLINE
without a
HOTLINE
Fluid Warming Set in place.
Ground Bond Test
Ground bond test must be tested according to methods and pass/fail
criteria described in IEC 60601-1:2005.
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...