S E C T I O N 3
•
I m p o r t a n t S a f e t y I n f o r m a t i o n
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
7
Continued from previous page
WARNINGS
WARNINGS
• Do not use the HOTLINE
®
Warmer in high-energy fields such
as: MRI, X-RAY, portable and mobile RF communications
equipment, and other such devices. The HOTLINE
®
Warmer
may act as a projectile in a strong magnetic field, cause image
artifacts, or not function as intended.
• Exposed conductor on MAINS power cord can cause an
electrocution hazard. Remove device from service if the MAINS
power cord has exposed wires.
• Grounding reliability can only be achieved when the MAINS
power cord is connected to a properly grounded receptacle.
Risk of electrical shock exists if the equipment is not connected
to a properly grounded receptacle.
• Do not mount the HOTLINE
®
Warmer more than 107cm (42")
above the floor. For convenience, 107cm (42") is indicated on
the HOTLINE
®
Warmer power cord by a black mark. Mounting
the HOTLINE
®
Warmer above 107cm (42") may result in
instability of the pole and tipping.
• Ensure that the HOTLINE
®
Warmer clamp is screwed tightly
onto the I.V. pole. Failure to securely mount the HOTLINE
®
Warmer onto the I.V. pole may cause the HOTLINE
®
Warmer to
slide down the I.V. pole.
• Do not use the HOTLINE
®
Warmer if equipment or Disposable
Set malfunction is evident.
• No user-serviceable parts. All service must be performed by
Smiths Medical or competent personnel.
• No modification of this equipment is allowed.
CAUTIONS
Malfunction, failure, or damage to the device may occur if these
cautions are not followed.
• Federal (U.S.A.) law restricts this device to sale by or on the
order of a physician.
• Do not autoclave or immerse any part of the HOTLINE
®
Warmer
in liquids, which may cause damage and improper functioning.
Continued
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...