Modes of Operation
The HOTLINE
Warmer operation is defined in the following modes:
• OFF Mode
• ON/Operating Mode
• Check Disposables Mode
• Add Recirculating Solution Mode
• Over Temperature Alarm Mode
The description of each mode includes a definition of the mode, activation
and/or monitoring of the mode, mode characteristics, and method to clear
the mode state.
OFF Mode
The power switch is in the OFF position (
a
) and the HOTLINE
Warmer is turned off.
ON/Operating Mode
The power switch is in the ON position (
b
) and the HOTLINE
Fluid
Warming Set has been properly installed.
Mode Characteristics
• The green Operating LED (
c
) illuminates.
• The reservoir temperature display will begin to increase.
• The recirculating solution path in the HOTLINE
Fluid
Warming Set will automatically prime.
Check Disposables Mode
The Check Disposables mode indicates a missing or improperly
installed HOTLINE
Fluid Warming Set.
Mode characteristics
• The green Operating LED on the Display Panel turns off.
•
he red Check Disposables LED (
d
) on the Display Panel
illuminates.
• The audible alarm sounds and repeats approximately every two
seconds.
• The recirculating solution stops circulating.
a
b
c
d
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...