S E C T I O N 9
•
T e s t i n g
26
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
To verify the recirculating solution temperature, you will need the
following:
• TEMP CHECK (HLTA-40)
• HOTLINE
Warmer
• HOTLINE
Fluid Warming Set
To Verify the Recirculating Solution Temperature:
1
Plug the
HOTLINE
Warmer into a power outlet.
2
Place the TEMP CHECK on the top right corner of the
HOTLINE
Warmer and plug it into the socket on the right side of
the HOTLINE
Warmer.
3
Plug the Twin-Tube Connector on the
HOTLINE
Fluid Warming
Set into the socket on the right side of the TEMP CHECK.
4
Remove the black label from the auxiliary outlet on the back of the
HOTLINE
and plug in the TEMP CHECK power cord.
Note: The auxiliary outlet is for use only with Smiths Medical
accessories.
5
Turn ON the
HOTLINE
Warmer. Allow 15 minutes for the
temperature to stabilize.
6
If the TEMP CHECK display indicates a temperature between
39
°
C and 40
°
C, and the
HOTLINE
Warmer display equals the
TEMP CHECK display, recirculating solution verification is
complete. Refer to the TEMP CHECK Manual for OVERTEMP
ALARM verification.
7
If the TEMP CHECK display does not indicate a temperature
between 39
°
C and 40
°
C, refer to the TEMP CHECK Manual for
calibration instructions.
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...