TECHNICAL DATA
The documentation includes a description of the rated characteristics of the electrical connection, all the intake and
outlet connections and the ambient conditions for which the appliance was designed. For full details, also refer to the
“INSTALLATION” section (ref. point 5.4.2 of IEC61010-1:2001).
For “non electric” intakes, the pressure and flow ranges are provided (ref. point 5.4.2 IEC61010-2-040:2005).
WATER SUPPLY REQUIRED
Type of water required
1 - Mains (required)
2 - Demineralised
(recommended)
Pressure
200 kPa – 500 kPa (2 bar – 5 bar)
Type of threaded connection
3 / 4 ”
Required flow [min - max]
2 - 12 litres / min
Mains water hardness [max]
42 °f
(for models equipped with a steam condenser, the use of purified water with hardness
below 15°f is recommended).
Fe2+ / Fe3+ [max]
0.5 ppm
Cold water temperature [max]
35°C
Demineralised water conductivity [max]
30μS/cm
Limits depending on microbial
contamination
Minimum microbiological quality required: “potable water" (ref. limits
in accordance with Italian Legislative Decree 31/2001)
DIMENSIONS and WEIGHT
Standard External
[Height x Width x
Depth]
850 x 600 x 600
Standard External - aquastop accessory
[Height x Width x Depth] - overall height is
increased by
+7mm
857
x 600 x 600
Standard External - PAD1 accessory
[Height x Width x Depth] - overall depth is
increased by
+110mm
850
x 600 x
710
Standard External - peristaltic pump P3
accessory
[Height x Width x Depth] -
overall depth is increased by
+60mm
850
x 600 x
660
Net weight
72 kg
Weight of app packaging
90 kg
MATERIALS
Washing chamber
AISI 316L
External surface