5.3
“
SET“ PARAMETERS
Set
on the display.
To access this item press the
Start/Stop
button.
The
Set
item contains a number of setting parameters.
Set
PARAMETER
DESCRIPTION
POSSIBLE USER
ACTION
POSSIBLE
SUPERUSER
ACTION
-1
Detergent pump P1 dispensing rate.
Setting in ml/litre, with maximum threshold of 10 ml/litre.
The default value of this parameter is
5ml/litre
.
Maximum recommended quantity with Smeg detergents:
10ml/litre.
Increase and decrease with the
Selection
buttons.
Display Only
Display and
Modification
-2
Detergent pump P2 dispensing rate.
Setting in ml/litre, with maximum threshold of 20 ml/litre.
The default value of this parameter is 3ml/litre.
Maximum recommended quantity with Smeg detergents:
10ml/litre.
Increase and decrease with the
Selection
buttons.
Display Only
Display and
Modification
-3
Detergent pump P3 dispensing rate.
Setting in ml/litre, with maximum threshold of 20 ml/litre.
The default value of this parameter is 10ml/litre.
Maximum recommended quantity with Smeg detergents:
15ml/litre.
Increase and decrease with the
Selection
buttons.
Display Only
Display and
Modification
-4
Water hardness setting
Set the hardness in °f, in steps of 5. Range: 5-60°f. The
default setting is 40°f.
Warning: Make sure that the value is actually the same as
the hardness of the water used.
Display Only
Display and
Modification
-5
Automatic opening at end of cycle
The default setting on
GW
appliances is “
On
”, but automatic
door opening can be disabled by changing the parameter to
“
OF
”.
If the option is enabled, the device door is opened
automatically at the end of the cycle.
Display Only
Display and
Modification
-6
Printer language
English by default, the language can be chosen from the
various options available
Display Only
Display and
Modification
German
French
Spanish
Italian
English
dE
Fr
ES
It
En
-7
Demineralised water
The use of demineralised water can be activated or
deactivated depending on whether or not the relative
connection has been made (Activated:
dn
, Deactivated: --)
Display Only
Display and
Modification
-8
Residual filter hours
The parameter indicates, to the nearest ten, the number of
absolute filter hours left before the maintenance message
appears. The initial default setting is 500. (e.g. the display
shows “-8:50”). The value can be reset by an authorised
engineer.
Display Only
Display Only