ACCESSING AND REUSING THE DEVICE AFTER AN INCOMPLETE CYCLE
The instructions relating to the device’s safety in the event of an incomplete operating cycle are provided in
compliance with points 5.4.4.g and 13.1.102 of IEC61010-2-040:2005.
WARNING
Warning applicable if the device is used for processing of contaminated materials.
If a disinfection cycle is interrupted (by the user or due to an alarm generated by the appliance
itself): take care when handling the elements and the load in general in the washing chamber.
The load and the internal parts of the appliance might be biologically contaminated/infected.
Before handling tools or performing any maintenance: Perform a complete disinfection cycle or,
if you cannot perform a complete Thermo-disinfection cycle, handle tools with caution (using
protection devices required for the handling of infected instruments, eg. gloves, lab coat).
DANGER, HOT SURFACES
The appliance performs a thermal disinfection cycle using water at high temperatures (up to 93°C)
and detergents. If, in the event of a failure, there is water in the chamber when the door is open:
avoid contact with the skin, burning hazard and risk of irritation due to the toxicity of the
chemicals used.
Never touch the heating elements inside the chamber.
Contact Smeg authorised technical staff in the event of a failure.
fig. 3 – View of inside of chamber, detail of heating elements.
3.2
DOOR OPENING
The following is a brief preview of the procedure for opening the door of the device for easier access to the chamber.
The points which follow provide a complete description of all the controls.
The appliance is equipped with an
automatic door locking system
.
Do not force it open; proceed as follows:
Connect the appliance to the electricity supply.
Hold down the button with the
On/Off
symbol for 2 seconds to switch the appliance on.
Press the
Door opening
button: the appliance gives a
beep
when it receives the signal. Release the button
and the door will open automatically after a few moments.
Do not touch
the heating
elements