HRX-OM-W002-A
Chapter 8 Control, Inspection, Exchange and Cleaning
HRR Series
8.4 Replacement of consumables
8-15
4.
The connection fitting of the DI filter and the tube is connected by a fastener. O ring is used
for the connection fitting. After removing the fastener, remove the connection fitting. Be
careful not to apply force to the tube at this time. Also, please be careful not to damage the O
ring.
Both the IN side and the OUT side have the same shape. Remove the connection fitting on
both sides and remove the DI filter.
5.
Install the DI filter in the reverse procedure that you removed.
There is no flow direction in the DI filter.
Connect the DI filter IN side tube to the bottom of the DI filter and the DI filter OUT side tube to the
upper side.Conversely, if the connection is made, the air in the DI filter will not come out and the
function of the DI filter may not work properly.
6.
Fix DI filter with DI filter fixing band.
7.
Install the filter panel. (Recommended tightening torque
:
1.5N
・
m)
Connection
fitting
fastener
DI Filter
①
②
O ring