Steam Cleaning
This product may be steam cleaned to eliminate bacteria and/or bed bug infestations, if desired.
When applying steam, avoid areas such as electrical connections and components including scale
meters, scale system components, hand and/or footboard controls that may be damaged by being
exposed to excessive moisture and/or heat.
Rinse all surfaces with a clean cloth dampened with fresh water to remove chemical and organic
residue.
After cleaning, all surfaces are to be wiped with a clean, dry cloth to remove any moisture or
residue.
Scale System Components
All surfaces of the scale meter and the coiled cord are to be wiped using a coarse cloth, dampened
with the disinfectant solution, prepared as directed by the manufacturer’s recommendations, to
remove organic material and visible soil.
Avoid excessive moisture to prevent damage.
Allow the scale meter and coiled cord to remain wet with disinfectant solution for the
manufacturer’s recommended contact time.
Rinse all surfaces of the scale meter with a clean cloth dampened with fresh water to remove
chemical and organic residue.
After cleaning, all surfaces are to be wiped with a clean, dry cloth to remove any moisture or
residue.
Bed Electronics and Hand Controls
Electronic components (Electrical Control Module, electrical connections and motor box covers
and hand controls) are to be wiped using a coarse cloth, dampened with the disinfectant solution,
prepared as directed by the manufacturer’s recommendations, to remove organic material and
visible soil.
Avoid excessive moisture to prevent damage.
Allow the electrical components and hand control to remain wet with disinfectant solution for the
manufacturer’s recommended contact time.
Rinse all electrical components with a clean, dry cloth to remove any moisture or residue.