50
Table 201 Guidance and Manufacturer’s Declaration – Emissions
Equipment and Systems that are NOT Life-Supporting
The Bari Rehab Platform 3™ is intended for use in the electromagnetic environment specified
below. The customer, or user, of the bed should ensure that it is used in such an environment.
Emissions Test
Compliance Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 2
The bed must emit electromagnetic energy in
order to perform its intended function. Nearby
electronic equipment may be affected.
RF Emissions
CISPR 11
Class B
The bed and RSS are suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonics Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/ Flicker
IEC 61000-3-3
Complies