54
General Safety
WARNING:
The bed is to be used in accordance with each facility’s policies and
procedures.
CAUTION:
Possible fire hazard when used with oxygen administering equipment
other than the nasal, mask or 1/2 bed length type tent. Oxygen tent should not
extend below the mattress support level. Lock hand control at foot of bed when
using oxygen administering equipment.
WARNING:
Electric shock may occur when plugging the bed into the wall outlet.
Use ONLY grounded or hospital grade outlets.
DO NOT use the bed if the power cord is cut, frayed or loosely connected to the
bed.
Pinch Point
Watch for pinch points on the bed, which are indicated by the pinch point label.
HAZARD:
Pay close attention to pinch points on the bed to avoid injury to the
patient and/or care provider.
Pinch points have minimal clearances between two moving parts that fail to maintain a clearance
of
≤
8mm (0.315”) or
≥
25mm (0.984”).
Pinch points are present on both sides of the bed between the bed and
the ground.
Pinch points are present between the mattress stop bar and footboard.