General Patient Room Cleaning/Disinfecting
Personal Protective Equipment should always be used as directed by the Material Safety Data
Sheet for the disinfectant.
Prepare the disinfectant according to the manufacturer’s recommendations.
Prepare a separate bucket of warm, fresh water to be used for rinsing the equipment after
cleaning/disinfecting procedures are completed as instructed.
Inspect sleep surface for attachment straps that may be securing the sleep surface to the bed top
deck and disconnect attachment straps.
For any active therapy based sleep surface, disconnect hoses from the control unit to prevent
damage to control unit or hose connections.
Remove sleep surface from the bed to gain full access to bed top deck and internal surfaces of
headboards, footboards and side rails.
Raise all bed functions to the highest position for easy access to all surfaces.
Bed to highest position
Head articulation to highest position
Foot articulation to highest position
Disconnect the bed from all electrical power to avoid electrical shock.
All bed frame surfaces are to be wiped using a coarse cloth, dampened with the disinfectant
solution, prepared as directed by the manufacturer’s recommendations, to remove organic material
and visible soil.
When cleaning components, such as hand and/or foot controls and areas where there are electrical
connections and components, avoid excessive moisture to prevent damage.
Allow all bed surfaces to remain wet with disinfectant solution for the manufacturer’s
recommended contact time.
Rinse all surfaces with a clean cloth dampened with fresh water to remove chemical and organic
residue.
After cleaning, all surfaces are to be wiped with a clean, dry cloth to remove any moisture or
residue.
NOTE:
Additional cleaning may be completed, as desired, with a mild soap solution and/or
household cleaning products. Avoid using harsh chemicals, such as acetone or paint thinner, as
they will damage the finish of the paint.