Fault diagnosis for DAC Professional
Fault 1: Vacuum system
CONFIGURATION OFCOMPONENT GROUPS IN THE AUTOCLAVE
ERROR 1 VACUUM UNIT
K
C
VV1
F
PV
VV2
VDA2
VQ
SLW
SQ1
PQ
VB
VDA1
H1
F
F
DAC Professional Configuration of the component
groups in the autoclave
(TOP VIEW)
FRONT
K
Item
Bezeichnung
Description
State of valve
when no
current is
applied (dead)
ACOUT
(power
output)
C
Kühler
cooler
4
F
Filter
filter
H 1
Dampferzeuger steam
generator
1
K
Kessel
vessel
P Q
Pumpe aqua dest
pump for distilled water
5
P V
Vakuumpumpe
vacuum pump
3
S LW
Sensor Leitwert
sensor for conductivity
S Q1
Strömungswächter
flow sensor for dist. water
V
Ventil (Magnet~)
valve (solenoid)
V
V
V
B
B
B
Belüftung
ventilation
open
8
V DA1
Druckablaß 1
pressure release 1
closed
7
V DA2
Druckablaß 2
pressure release 2
closed
9
V
V
V
Q
Q
Q
Speisewasser aqua dest
feed water (distilled)
closed
5
V
V
V
V
V
V
1
1
1
Vakuum 1
vacuum 1
closed
6
V
V
V
V
V
V
2
2
2
Vakuum 2
vacuum 2
closed
6
This overview shows a rough schematic of the spatial configuration of the
component groups, pipes, and hoses, in the DAC Professional. The following error
descriptions will refer to this overview to help you find the component groups to be
checked.
09/21/2006
15/61
Servicemanual DAC Professional
Содержание DAC PROFESSIONAL
Страница 1: ...b a PROFESSIONAL p j...
Страница 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Страница 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Страница 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Страница 64: ......
Страница 65: ...61 75 538 D 3472 D 3472 076 02 01 02 1 b...