Fault Diagnosis DAC Professional
Fault 9: Door open
FAULT 9: DOOR OPEN
Triggered by:
Door contact (digital input DIN1) was closed during a program
Causes/ Remedy
Operating fault/ Checks by personnel:
•
Press down door handle as far as possible
Appliance fault / Repair by technical service personnel
►
Door switch faulty
•
Door switch faulty / needs readjustment
•
Electrical connection to CPU-board interrupted
•
Digital input DIN1 on CPU-board faulty
Check in the Diagnosis-Program, "Digital Inputs: E 1“
if the door switch needs adjusting, see below.
Adjusting the door contact switch
•
In the diagnosis program "Digital Inputs“ select
"Input E1: Door contact“
•
Insert 8 mm spacer (M8 bolt or similar) between
lock housing and upper door beam as shown
•
Close door and push down door lock until contact is made
•
Loosen right fixing screw of door contact switch
•
Turn door contact switch around the left fixing
screw until door contact switch engages (Display
shows "closed", switching sound)
•
Tighten right adjusting screw in new position
09/21/2006