Fault-diagnosis DAC Professional
Fault 18: Sensor "nr" Input "nr"
Display message
Triggered by/Cause
Correction
Comments / Remarks
Fault18
Sensor: 1 Input 1
Temperature at analog input
AIN1
≤
0°C or
≥
T23
•
Temperature sensor faulty
Short-circuit/Open
•
Signal processing on CPU-
board faulty
•
Exchange
temperature
sensor
•
Exchange
CPU-
board
Until the sensor/CPU-board can be
exchanged, in the Service-program
set FLAG F28 to 1, AIN 2 then
takes over the function of
AIN 1
Fault18
Sensor: 2 Input 2
Temperature measured at
analog input
AIN2
≤
0°C or
≥
T23
•
Temperature sensor faulty
Short-circuit/Open
•
Signal processing on CPU-
board faulty
•
Exchange
temperature
sensor
•
Exchange
CPU-
board
Until the exchange of the
sensor/CPU-board, unplug the
defective sensor on input AIN2,
move the sensor on AIN1 to AIN2
and in the Service-Program set
F28 to 1
Fault18
Sensor: 3 Input 3
Pressure measured to analog
input AIN3
≤
0 mbar or
≥
P19
•
Pressure sensor faulty
•
Signal processing to CPU-
board faulty
•
+ 24V (voltage to pressure
sensor) at CPU-board faulty
•
Exchange
pressure sensor
•
Exchange
CPU-
board
Sequence of testing:
+24V supply (between GND and
+UB) measured at pressure
sensor
Signal voltage at pressure sensor
between GND and + Signal must
be around 2.5V (corresponds
1000mbar air pressure)
Fault18
Sensor: 4 Input 4
Temperature measured at
analog input AIN4
≤
0°C or
≥
T23
•
Temperature sensor faulty
Short-circuit/Open
•
Signal processing on CPU-
board faulty
•
Exchange
temperature
sensor
•
Exchange
CPU-
board
Until exchange of the sensor/CPU-
board, in the Service program set
FLAG F5 to 0, this deactivates the
entire preheating including
malfunction reports - drying quality
will be reduced
Fault18
Sensor: 6 Input 6
Conductivity measured at
analog input AIN6
≥
L3
•
Check for short-
circuits of the
conductivity
sensor
•
Exchange
CPU-
board
Until exchange of the CPU-board,
in the Service program set FLAG
F3 to 0, this deactivates the entire
conductivity measurement
including malfunction reports -
temporarily no automatic checks of
water quality ,
09/21/2006
33/61
Servicemanual DAC Professional
Содержание DAC PROFESSIONAL
Страница 1: ...b a PROFESSIONAL p j...
Страница 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Страница 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Страница 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Страница 64: ......
Страница 65: ...61 75 538 D 3472 D 3472 076 02 01 02 1 b...