Fault-diagnosis DAC Professional
Fault 32: Power loss/Sterilize sterile-filter
FAULT 32: Power loss / Sterilize sterile filter
Triggered by :
After the start of a program there is a loss of the power supply. The fault is reported as soon as the electric power
supply is resumed:
If there has been a power loss then on restarting, there will be a reminder on the display to sterilize the sterile filter,
since this may have been damp and be contaminated with pathogens:
Remove sterile filter on the rear side of the autoclave
Sterilize the filter in the quick program
After completion of the program replace the sterilized filter
Causes/ Remedy
Operating fault/Check by personnel:
•
Manual program termination by switching off the autoclave from the mains supply
•
Check for mains power cut, and intact electrical supply installations
Appliance fault / Repair by technical service personnel
•
Power switch faulty
•
Main fuses F1, F2 (on front of autoclave beneath power switch) is faulty
•
Check internal wiring of the autoclave (in particular power supply to the CPU board)
•
Power supply on the CPU board is faulty
09/21/2006
41/61
Servicemanual DAC Professional