Operating Instructions CEREC MC XL Premium Package
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter inserts are completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait briefly until the filter inserts are soaked full and add a
corresponding amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see “Switching the unit ON and OFF [ → 32]”).
9. Switch the pump on (press the
"Pump"
button on the touch display)
to fill the water circuit.
10. Fill the water tank up again until the filter inserts are completely
immersed (up to the bottom edge of the cover thread).
5.4.9
Switching the unit ON and OFF
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
Within the machine, grease depots are included for lubricating
components that can cause error messages at low temperatures.
✔
Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is absolutely
dry (for at least one hour)
The unit is dry and can be put into operation.
NOTICE
Do not adjust the line voltage
The unit automatically adjusts to the line voltage.