63 98 536 D 3439
54
D 3439
.
201.06.10
.
02 10.2017
7 Service
Dentsply Sirona
7.3 Changing the water
Operating Instructions CEREC MC XL Premium Package
7.3.2.1
Procedure for all materials except for base metals
To change the water, proceed as follows:
✔
The unit is switched on.
✔
No machining / scan process is running.
1. Pull out the water tank at the front of the unit.
2. Open the drain opening (D).
3. Empty two thirds of the water from the tank.
4. Close the drain opening (D).
5. Shake the tank vigorously.
6. Open the drain opening (D).
7. Drain the rest of the grinding water.
8. Close the drain opening (D).
9. Turn the tank cap (B) counter-clockwise and take it off.
10. Add approx. 75 ml of DENTATEC to the tank.
11. Fill the tank with water until the filter inserts (A) are completely
immersed (up to the bottom edge of the cover thread, approx. 3
liters).
12. Wait briefly until the filter inserts (A) are soaked full and add a
corresponding amount of water.
13. Close the water tank by tightening the tank cap (B) clockwise by
hand. Do not use the tank cap opener for this.
14. Push the water tank back into the housing.
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not
permitted.
Do not use any cleaning agents.
Содержание CEREC MC XL Premium Package
Страница 77: ......