63 98 536 D 3439
64
D 3439
.
201.06.10
.
02 10.2017
7 Service
Dentsply Sirona
7.8 Changing the filter
Operating Instructions CEREC MC XL Premium Package
3. Open the clamping ring and lay it down.
Disposing of filter cartridges and ceramic sludge
1. Release the filter cartridges by tilting them slightly. The springs
remain attached.
Tip: If a spring accidentally comes loose, you can reattach it by
screwing it in counter-clockwise.
2. Dispose of the filter cartridges with domestic waste.
3. Slowly pour the water out of the container into the outlet so that the
ceramic sludge remains in the container.
4. Take the plastic bag out of the container and dispose of it with
domestic waste.
Preparing the external water tank
1. Insert the plastic bag into the external water tank in such a way that
it protrudes by around 3-4 cm.
2. Ensure that it protrudes evenly around the edge of the container.