6 Operation
Dentsply Sirona
6.3 Machining process
Operating Instructions CEREC MC XL Premium Package
6.3.4
Ending the machining processes
1. When the machining process has been completed, open the
processing chamber door.
2. Remove the restoration.
3. Loosen the ball pressure screw.
4. Remove the remainder of the ceramic block. When removing the
remaining block from blocks with 6 mm diameter block holders, make
sure that the adapter sleeve remains in the machine.
5. Close the processing chamber door.
NOTICE
Ending the machining processes
You can terminate the machining process at any time by clicking the
"Stop"
button on the PC or on the touch display.
WARNING
Risk of injury on the remainder of the ceramic block
The remaining portion of the ceramic block may have sharp edges
(e.g. A) that could injure you if it is not removed carefully.
Always grasp the remainder of the ceramic block by its metal holder.
CAUTION
Do not use defective milling and grinding results!
Machining results must be judged by the user (dentist or dental
technician) and must not be used if defects are detected!
NOTICE
If you have not used the milling and grinding unit for a rather long time,
we recommend you should switch it off and then open the processing
chamber door so that the processing chamber can dry out.