37
Dentsply Sirona
6 Operation
Operating Instructions CEREC MC XL Premium Package
6.1 Configuration (CEREC MC XL Premium Package)
båÖäáëÜ
● Bar code reader
– The barcode reader which is available as standard can be
deactivated via the corresponding check box on the PC or touch
display under
"Edit Device Settings"
.
● External tank
– If the external water tank is connected and the check mark has
been placed, you will not be reminded to change the water until
a later point in time.
– You can deactivate the external water tank by switching off the
check mark in front of
"External Water Tank"
on the PC screen
or the
"External Tank"
on the touch display under
"Edit Device
Settings"
.
● Extraction Unit
- If a suction device is connected, and the check mark is ticked, the
dry milling process is automatically started when processing
zirconium oxide. To deactivate dry milling, you can remove the check
mark once again.
Deactivation of an instrument set (only for units with four motors)
It is likely that you will have to deactivate an instrument set. For example,
if a defective instrument could not yet be replaced or if a motor is defective
or cannot be calibrated.
In all of these cases, you can deactivate sets 1 and 2 separately. A
deactivated set is ignored during processing, calibration, etc.
➢ You can deactivate an instrument set by removing the check mark in
front of the instrument set in the software or by deactivating the
instrument set on the touch display under
"Edit Device Settings"
.
NOTICE
Milling instruments must only be used in Set 1.
NOTICE
Risk of collision
The restoration may become damaged if longer instruments are present
in the deactivated instrument set than in the active set.
➢ Ensure that the instruments installed in the deactivated instrument
set are not longer than those in the active set.