63 98 536 D 3439
36
D 3439
.
201.06.10
.
02 10.2017
6 Operation
Dentsply Sirona
6.1 Configuration (CEREC MC XL Premium Package)
Operating Instructions CEREC MC XL Premium Package
6
Operation
6.1
Configuration (CEREC MC XL Premium
Package)
In the
"Devices"
area of the
"CEREC SW"
software, various settings can
be subsequently modified.
1. Click the
"Configuration"
button in the system menu.
2. Click on the
"Devices"
button.
3. Click on the unit that you wish to configure.
CEREC MC XL Premium Package
You can subsequently edit the following settings via the menu item
"MC
XL"
:
● Name
● Connection settings
– Retrieve IP settings automatically
– Specify IP settings manually
● Manual block fixing
– If you use manual block fixing, a check mark must be placed in
front of
"Manual Block Chuck"
.
– Models can only be milled using manual block fixation.
● Second motor set
– The check mark must be placed in front of
"Two Bur Sets"
.
– You can deactivate instrument sets individually. A deactivated
instrument set is ignored during processing, calibration, etc.
NOTICE! The restoration may become damaged if longer
instruments are present in the deactivated instrument set than in
the active set. Ensure that the instruments installed in the
deactivated instrument set are not longer than those in the active
set.
CAUTION
Risk of injury on calibration pins/instruments
If you reach into the processing chamber (e.g.: when inserting/removing
a ceramic block, changing instruments, or inserting/removing a
calibration phantom), you may injure your hand on the calibration pins/
instruments.
Be careful not to brush against the calibration pins/instruments with your
hand.
Always insert your hand in the processing chamber underneath the
calibration pins/instruments.
Содержание CEREC MC XL Premium Package
Страница 77: ......