6 Operation
Dentsply Sirona
6.2 Calibrating the unit
Operating Instructions CEREC MC XL Premium Package
6.2
Calibrating the unit
Unit calibrated ex works
The unit is calibrated at the factory. No additional calibration is required
during initial startup. Proceed as described below when performing a
subsequent calibration.
Preparing a calibration
1. Take the calibration pins and calibration phantom out of the drawer of
the unit.
2. In the software, go to the system menu and click on the
"Configuration"
button.
3. Click the
"Devices"
button.
4. Click on the unit that you wish to calibrate.
5. Click the
"Calibrate"
step.
If two instrument sets are active:
A dialog box then opens where you can select the instrument set
to be calibrated or the two instrument sets to be calibrated
consecutively. The date of the last calibration is also displayed.
This information also appears on the touch display.
6. If necessary, select the desired instrument set.
You can also select the desired instrument set on the touch display.
7. Click the
"Start"
button on the PC or on the touch display.
The unit moves to the insertion position for the calibration tools.
A dialog window prompts you to insert the calibration pins and
calibration phantom and to close the processing chamber door
again.
NOTICE
Use only the supplied calibration tools
Use only the supplied calibration pins and the corresponding calibration
phantom when calibrating the unit.
NOTICE
Incorrect machining result
If a unit is not calibrated, the machining result may be incorrect.