Operation
TP17 / TPM
- 36 -
© SIKA • Ba_TP17-TPM_en • 01/2016
Setting values for „SLor“ and „SLoF“
Calibrator type
Heating gradient
1)
„SLor“
Cooling gradient
2)
„SLoF“
Heating/Cooling:
TP 17 165 S
< 7 °C/min
< 5 °C/min
TP M 165 S / TP 17 166 /TP 17 200
-
with silicone oil 20CS
-
with distilled water
-
as dry block
-
as infrared calibrator
-
as surface calibrator
< 3 °C/min
< 5 °C/min
< 3 °C/min
< 3 °C/min
< 3 °C/min
< 6 °C/min
< 4 °C/min
< 4 °C/min
< 4 °C/min
< 4 °C/min
Heating:
TP 17450 S, TP 17650 S
< 35 °C/min
max. … 300 °C
300 °C … 100 °C
< 10 °C/min
< 5 °C/min
TP 17450 S
-
as infrared calibrator
-
as surface calibrator
< 35 °C/min
< 30 °C/min
max. … 300 °C
300 °C … 100 °C
max. … 300 °C
300 °C … 100 °C
< 10 °C/min
< 5 °C/min
< 10 °C/min
< 5 °C/min
TP M 225 S / TP M 255
-
with silicone oil 20CS
-
with distilled water
-
as dry block
-
as infrared calibrator
-
as surface calibrator
< 22 °C/min
< 12 °C/min
< 12 °C/min
< 12 °C/min
< 10 °C/min
200 °C … 50 °C
50 °C … 30 °C
90 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
200 °C … 50 °C
50 °C … 30 °C
< 4 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
< 2 °C/min
< 0,5 °C/min
1)
Heating gradient „SLor“:
The heating gradient "SLor" is active if the reference temperature is lower than the set
temperature. Each calibrator type has a max. heating capacity, meaning that only settings <
than this heating capacity are reasonable and extend the time until the set temperature is
achieved.
2)
Cooling gradient „SLoF“:
The cooling gradient "SLor" is active if the reference temperature is higher than the set
temperature. Only settings below the cooling capacity of the calibrator have an effect on the
cooling gradients.
Duration time „dur.t“:
The duration time "dur.t" is active if the set temperature SP1 has been achieved.
Subsequently, the calibrator / micro bath automatically switches to set temperature SP2.