Technical data
TP17 / TPM
- 52 -
© SIKA • Ba_TP17-TPM_en • 01/2016
10.3
Characteristics TP 17 … S
Series
Characteristics
TP
17 165 S
TP
17 166 S
TP
17 200 S
TP
17 450 S
TP
17 650 S
Calibrator
Temperature range:
-
Dry block
-35…165 °C -35…165 °C -55…200 °C T
R
…450 °C T
R
…650 °C
-
Infrared
-/-
-35…165 °C
-/-
T
R
…450 °C
-/-
-
Surface
-/-
-25…150 °C
-/-
T
R
…400 °C
-/-
Accuracy:
-
Dry block
±
0.2 °C
±
0.2 °C
±
0.2 °C
±
0.3 °C
±
0.4 °C
-
Infrared
-/-
±
0.5 °C
-/-
±
0.5 °C
-/-
-
Surface
-/-
±
1.0 °C
-/-
±
1.0 °C
-/-
Emission degree:
-
Infrared
-/-
0.9994
-/-
0.9994
-/-
Stability:
-
Dry block
±
0.05 °C
-
Infrared
±
0.05 °C
-
Surface
±
0.2 °C
Display range:
-50…165 °C -50…165 °C -60…200 °C 0…450 °C
0…650 °C
Electrical characteristics
Supply voltage:
-
100…240 V
AC
50/60 Hz
-/-
50/60 Hz
-
230/240 V
AC
-/-
-/-
-/-
50/60 Hz
-
100…115 V
AC
(optional)
-/-
-/-
-/-
-/-
50/60 Hz
Power consumption
375 W
555 W
2000 W
1000 W
Fuse:
-
100…240 V
AC
6.3 A slow
-/-
10.0 A slow
-
230/240 V
AC
-/-
-/-
-/-
10.0 A slow 6.3 A slow
-
100…115 V
AC
(optional)
-/-
-/-
-/-
-/-
10.0 A slow
Serial interface:
-
Serial interface type
RS 485 optoisolated
-
Communication protocol
MODBUS RTU (JBUS)
-
Serial transfer speed
9600 Baud
Process variables
Test specimen holder:
-
Bore
∅
28 mm
∅
60 mm
∅
28 mm
∅
60 mm
∅
28 mm
-
Depth
150 mm
-
Measurement zone
110…150 mm • Infrared: 110 mm • Surface
Housing dimensions:
-
Width
210 mm
150 mm
-
Height
380+50 mm
330+70 mm
-
Depth
300 mm
270 mm
Weight
~10.0 kg
~10.0 kg
~12.5 kg
~7.5 kg
Sleeves for smaller
∅
(in 0.5 mm steps)
1.5…25 mm 1.5…55 mm 1.5…25 mm 1.5…55 mm 1.5…25 mm