TP17 / TPM
Commissioning and operation
Technical changes reserved
- 15 -
4.2
Electrical connection
The electrical connections are made with the accompanying mains connecting cable.
DANGER! Risk of death due to electric current!
The TP17 / TPM is operated at mains voltages up to 240 V
AC
. Contact with the mains voltage
can result in serious or fatal injuries.
Switch off the TP17 / TPM and remove the mains connecting cable before you start
any work on live parts.
Check the following points before you connect the calibrator:
Operate the machine only with the approved supply voltage (
§ 10). Ensure that the
mains voltage is the same as that specified on the type plate.
Please follow the local regulations of the energy supplier.
Connect the calibrator only to a properly installed and earthed 3-pole outlet for Schuko
plugs.
Do not use any extension cables or adapter plugs.
IMPORTANT! Mains connecting cable!
The mains connecting cable may only be replaced by an equivalent cable.
Use only original cables from SIKA or approved cables of the same type with the
correct design as replacements (
§ 10.1…10.4 "Electrical characteristics").
Connect TP17 / TPM:
Connect the mains connecting cable with the connector plug of the TP17 / TPM.
Insert the plug of the mains connecting cable in a suitable outlet.