Technical data

 

TP17 / TPM  

- 54 - 

 

© SIKA  •  Ba_TP17-TPM_en  •  01/2016  

10.5

 

Heating and cooling periods 

 

The following measurement conditions apply to heating and cooling times:  

 

All times are elapsed times without settling time. 

 

The measurements were carried out at a room temperature of approx. 23 °C 

 

Without cap, without cover (surface function and TP M  … S series). 

 

Regulation is based on the internal reference sensor (except for surface). 

 

Dry block (series TP 17 … , series TP M … S): 

 

Reference sensor Pt100, Ø 6 mm, stainless steel casing 

 

TP M … S:  Reference sensor located 10 mm away from middle at a depth of 155 mm. 

 

TP 17 … : At full depth the reference sensor is located in the middle of the sleeve. 

 

Infrared (TP 17 166 S, TP 17 450 S, series TP M … S): 

 

All periods relate to a reference sensor Pt100 Ø3 mm, L = 300 mm, stainless steel 
casing.  

 

The reference sensor is immersed at a depth of 111 mm. 

 

Surface temperature (TP 17 166 S, TP 17 450 S, series TP M … S): 

 

Regulation is based on the external reference sensor (3 x 300 mm). 

 

Reference sensor Pt100 Ø3 mm, L = 150 mm, stainless steel casing. 

 

Reference sensor immersed at 51 mm depth and positioned below the upper edge. 

 

Micro bath (series TP M … S): 

 

Reference sensor Pt100, Ø 6 mm, stainless steel casing. 

 

The reference sensor is located 5 mm above the mesh insert in the middle of the tank. 

 

10.5.1

 

TP 17 165; TP 17 166, TP 17 200 (dry block) 

 

 

Содержание TP17 165 S

Страница 1: ...manual for future reference If the device is resold please provide the operating manual along with it SIKA Ba_TP17 TPM_en 01 2016 Temperature calibrators Series TP17 TPM Operating manual Translation...

Страница 2: ...calibrator 17 4 3 2 Calibrator for infrared thermometer 18 4 3 3 Calibrator for surface temperature sensors 19 4 3 4 Micro bath calibrator 20 4 4 Switching on cool down and switching off 24 4 5 Operat...

Страница 3: ...P 17 S 52 10 4 Characteristics series TP M S 53 10 5 Heating and cooling periods 54 10 5 1 TP 17 165 TP 17 166 TP 17 200 dry block 54 10 5 2 TP 17 450 TP 17 650 dry block 55 10 5 3 TP 17 166 S TP 17 4...

Страница 4: ...urrent This sign indicates dangers which could arise from handling of electric current WARNING CAUTION Risk of injury This sign indicates dangers that cause personal injuries that can lead to health d...

Страница 5: ...ents can be directly connected and checked Versions The series TP17 TPM includes the following calibrator micro bath types Metal block calibrators Micro baths TP 17 165 c h TP 17 165 S c h TP 17 450 h...

Страница 6: ...ure calibrators are delivered in special protective packaging Save the packaging for returning the instrument safely to the manufacturer for recalibration or repair Scope of delivery and accessories i...

Страница 7: ...elow 10 C Allow the calibrator to cool before you remove the test specimen clean the tank change the adapter sleeve or the calibration insert or switch off the machine CAUTION Material damage The open...

Страница 8: ...ansferred to our customers our warranty is not valid Qualified personnel The personnel who are charged for the installation operation and maintenance of the TP17 TPM must hold a relevant qualification...

Страница 9: ...losive atmosphere Remove all the easily flammable media from the vicinity of the calibrator Ensure that the calibrator cannot come in contact with easily flammable or explosive media Operate the calib...

Страница 10: ...oil recommended here Always read the safety data sheet supplied with the silicone oil before using it Always ensure adequate ventilation when working with silicone oil since hazardous substances can...

Страница 11: ...sing holds a heat insulated calibration block with heating or cooling elements and an integrated sensor for the reference temperature In the case of the dry block and multifunction calibrator this is...

Страница 12: ...the preparations have been carried out the calibrator can be switched on 4 4 The calibrator heats or cools the metal block or calibration liquid to the set temperature As soon as this temperature has...

Страница 13: ...ts for operation with a USB converter are IBM compatible PC installed operating system Windows 7 Windows 8 or Windows 10 a free USB port USB 1 1 or USB 2 0 A network configuration allows the connectio...

Страница 14: ...t conditions p 50 4 1 Operating conditions Select a safe installation site for commissioning the machine Installation site and operating position Only suitable for indoor use do not use outdoors Opera...

Страница 15: ...nly with the approved supply voltage 10 Ensure that the mains voltage is the same as that specified on the type plate Please follow the local regulations of the energy supplier Connect the calibrator...

Страница 16: ...imen clean the tank change the adapter sleeve or the calibration insert or switch off the machine IMPORTANT Empty the tank and clean it With micro bath calibrators TPM the tank must be emptied and cle...

Страница 17: ...r of the test specimen The measurement element of the test specimen must be located in the homogeneous temperature zone of the adapter sleeve Please also note the Details of the DKD technical committe...

Страница 18: ...must be smaller than the inner diameter and must not touch the wall of the infrared insert Inserting The infrared insert is inserted into the metal block with the aid of the sleeve exchange tool Remo...

Страница 19: ...surface insert from getting stuck in the metal block IMPORTANT NOTICE The calibration of surface temperature sensors is very difficult and not uncontroversial The temperature sensor which are placed...

Страница 20: ...e of the calibration liquid during transport CAUTION Safety valve The transport cover is equipped with a safety valve which is activated once the pressure reaches 1 5 bar This can result in hot steam...

Страница 21: ...he tub insert is placed into the tank with the help of the sleeve exchange tool Just like the tank the tub insert can be closed with the associated cover Both threaded covers are leak proof so the cal...

Страница 22: ...Exchange the worn magnet stirrer Exchange dirty smeary calibration fluid 4 3 4 2 Information on filling amounts CAUTION Risk of incorrect measurement or material damage Do not exceed the maximum fill...

Страница 23: ...nsion displacement by sensors and level rise due to stirring Screw on the cover of the tank tub insert Insert the tub insert in the tank using the sleeve exchange tool only for tub insert Place the ma...

Страница 24: ...During this process the calibrator has not yet reached the required insulation voltage for protection class I For drying the heating elements heat the calibrator for at least 15 min to 120 C Cool dow...

Страница 25: ...es the calibrator or the test specimen may be damaged Switch off the calibrator only when the metal block or tank has reached room temperature Check whether the metal block tank has cooled down to roo...

Страница 26: ...ro bath is heating If LED OUT 1 is not lit the calibrator or micro bath is not heating 9a LED OUT 2 a Heating instrument Indicates the status of the output for the fan control If LED OUT 2 is lit the...

Страница 27: ...r 5 2 2 Display of temperature control The LED OUT 1 indicates the status of the temperature control If LED OUT 1 is lit the temperature control is active The heating or cooling is switched on If LED...

Страница 28: ...and operating duration In calibration mode you can check the heating or cooling capacity and the operating duration of the calibrator The current values of the calibrator appear on the lower display...

Страница 29: ...lay Press the key to increase the set temperature Press the key to decrease the set temperature Press the P key again to confirm the new set point NOTES Press the and key to raise and lower the value...

Страница 30: ...le temperature calibrator or micro bath NOTICE The S version provides certain additional functions e g storage of four different set temperatures or setting of the control parameters Main menu for mul...

Страница 31: ...ed Please connect this to the mini DIN socket and turn the selector switch to ext Ref Menu structure S versions As displayed by the menu structure it is possible to reach the group and parameter level...

Страница 32: ...rence temperature will continuously change and adjust to the room temperature without having to be further regulated Switching on the automatic control The control is switched off if the following dis...

Страница 33: ...t capacity PLEASE NOTE Press the and key to raise and lower the value by 0 1 respectively If the keys are held pressed for at least one second the value increases or decreases quickly and after two se...

Страница 34: ...r key to select one of the four set point memories SP1 SP2 SP3 and SP4 Press the P key to open the respective set point memory The selected set point memory e g SP 3 flashes on the upper display The c...

Страница 35: ...erature profile S version It is possible to carry out a gradient control yourself and thus determine the time in which the set temperature is reached The time can be shorter or longer than the time us...

Страница 36: ...or 22 C min 12 C min 12 C min 12 C min 10 C min 200 C 50 C 50 C 30 C 90 C 50 C 50 C 30 C 200 C 50 C 50 C 30 C 200 C 50 C 50 C 30 C 200 C 50 C 50 C 30 C 4 C min 0 5 C min 2 C min 0 5 C min 2 C min 0 5...

Страница 37: ...ime can be set in the parameter level rEG Pressing the P key for approx 5 sec the main menu opens The last selected function appears on the upper display LED SET flashes on the lower display Press the...

Страница 38: ...ey to decrease the heating gradient SLor Press the and key to raise and lower the value by 0 1 respectively If the keys are held pressed for at least one second the value increases or decreases quickl...

Страница 39: ...cooling gradient SLoF Press the key to decrease the cooling gradient SLoF Press the and key to raise and lower the value by 0 1 respectively If the keys are held pressed for at least one second the va...

Страница 40: ...on the lower display Press the key to increase the duration time dur t Press the key to decrease the duration time dur t Press the and key to raise and lower the value by 0 1 respectively If the keys...

Страница 41: ...specimen requires a short period of time until it absorbs the temperature of the metal block or liquid bath The internal references are set to normal when operating the micro bath the dry block and th...

Страница 42: ...finished Write down the reference temperature and the temperature of the specimen for your test protocol Set the calibrators gradually to the set values of the further test points and proceed as descr...

Страница 43: ...ence sensor is defective The controller switches off the power supply to the heating cartridge servicing required uuuu Measured temperature under the limit value of the internal reference sensor under...

Страница 44: ...injuries e g caustic burns or toxic contaminations decontamination measures disposal etc that can be attributed to insufficient cleaning of the measuring instrument Comply with the instructions below...

Страница 45: ...calibrators check the filled height of the calibration liquid When doing so note the Information on filling amounts 4 3 4 2 Annually Subject all the parts of the calibrator to a visual inspection for...

Страница 46: ...only one is defective Fit the fuse compartment back in place and connect the mains connecting cable 4 2 Electrical connection Should the fuses blow repeatedly there is probably a fault in the calibrat...

Страница 47: ...is too low can result in the temperature fuse getting triggered The calibrator is then rendered operationally disabled and must be sent to SIKA Ensure that the grille openings are always clear Measuri...

Страница 48: ...n in operation Touching hot parts can result in serious injuries Allow the calibrator to cool down before transport Carry the calibrator only at the handle Observe the following instructions for a saf...

Страница 49: ...n cool down and switching off Decommissioning Remove all connected sensors and devices Switch off the calibrator and disconnect the mains plug Empty the calibrator tank in case there is residual calib...

Страница 50: ...display green lower display Display for sensor break Electrical characteristics Degree of protection IP 20 Block temperature control via PID controller Controller outputs Heater control Cooler contro...

Страница 51: ...Supply voltage 100 240 VAC 50 60 Hz 50 60 Hz 230 240 VAC 50 60 Hz 100 115 VAC optional 50 60 Hz Power consumption 375 W 555 W 2000 W 1000 W Fuse 100 240 VAC 6 3 A slow 10 0 A slow 230 240 VAC 10 0 A s...

Страница 52: ...trical characteristics Supply voltage 100 240 VAC 50 60 Hz 50 60 Hz 230 240 VAC 50 60 Hz 100 115 VAC optional 50 60 Hz Power consumption 375 W 555 W 2000 W 1000 W Fuse 100 240 VAC 6 3 A slow 10 0 A sl...

Страница 53: ...Display range 50 165 C 0 225 C 0 255 C Electrical characteristics Supply voltage 100 240 VAC 50 60 Hz 50 60 Hz 50 60 Hz 230 240 VAC 50 60 Hz 50 60 Hz 100 115 VAC optional 50 60 Hz 50 60 Hz Power cons...

Страница 54: ...e at a depth of 155 mm TP 17 At full depth the reference sensor is located in the middle of the sleeve Infrared TP 17 166 S TP 17 450 S series TP M S All periods relate to a reference sensor Pt100 3 m...

Страница 55: ...TP17 TPM Technical data Technical changes reserved 55 10 5 2 TP 17 450 TP 17 650 dry block 10 5 3 TP 17 166 S TP 17 450 S infrared surface...

Страница 56: ...Technical data TP17 TPM 56 SIKA Ba_TP17 TPM_en 01 2016 10 5 4 TP M 165 S Dry block infrared and surface Water and silicone oil...

Страница 57: ...TP17 TPM Technical data Technical changes reserved 57 10 5 5 TP M 225 S TP M 255 S Dry block infrared and surface Water and silicone oil...

Страница 58: ...perature block calibrator If this is the case the calibration is only effective if the calibrator was used with a corresponding heat transfer medium With the calibration of thermometers with an outer...

Страница 59: ...TP17 TPM Guideline DAkkS DKD R 5 4 Appendix B Technical changes reserved 59 For your notes...

Страница 60: ...1 2016 Sensors and Measuring Instruments Flow Measuring Instruments Test and Calibration Instruments SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 D 34260 Kaufungen Germany 49 0 5605 803 0 49 0 5605 8...

Отзывы: