Operation
TP17 / TPM
- 32 -
© SIKA • Ba_TP17-TPM_en • 01/2016
5.4.1
Automatic control
For certain tasks it can be advantageous to switch off the control, e.g. to carry out settings at
the calibrator / micro bath.
Switching off automatic control:
Press the
P
key when in calibration mode for approx. 5 sec., the main menu opens.
The last selected function appears on the upper
display.
LED SET flashes on the lower display.
Press the
or
key until
OFF
appears.
Press the
P
key to confirm.
An alternating display of the current reference
temperature and
OFF
appears on the upper
display.
The current set temperature appears on the lower
display
TAKE NOTICE OF:
The controller is now switched off and the reference temperature will continuously
change and adjust to the room temperature without having to be further regulated.
Switching on the automatic control:
The control is switched off if the following display
appears:
An alternating display of the current reference
temperature and
OFF
appears on the upper display.
The current set temperature appears on the lower
display.
Switch the control back on by
pressing the
P
key for approx. 5 sec., the main
menu opens.
OFF
appears on the upper display.
LED SET flashes on the lower display.
Confirm switching on the controller by pressing
the arrow key until the desired operating mode is
displayed and confirm this with the
P
key.
PLEASE NOTE:
The control has been reactivated. The calibrator / micro bath is in calibration mode and
the set temperature is targeted.
Menu control OFF
Control OFF setting display
Control OFF setting display
OFF display