TP17 / TPM
Commissioning and operation
Technical changes reserved
- 25 -
IMPORTANT! Mains failure or separation from the mains!
If there is a mains failure, or if the main switch is turned off, or upon removal of the
mains connection ("EMERGENCY STOP"), the built-in fan does not blow any more cooling
air.
Sufficient thermal de-coupling between the metal block, the tank and the housing is
nonetheless guaranteed
Switching off:
CAUTION! Before switching off, pay attention to the block temperature!
The calibrator should first reach a low block temperature before switching off. When
switching off at high temperatures, the calibrator or the test specimen may be damaged.
Switch off the calibrator only when the metal block or tank has reached room
temperature.
Check whether the metal block / tank has cooled down to room temperature.
Switch off the calibrator via the main switch.
Disconnect the calibrator from AC power if no further testing is required.
Unplug the power cord from the wall outlet.
Clean the calibrator after use (
§ 8.2).
4.5
Operating modes
During operation, there are three operating modes:
Calibration mode:
This is the normal operating mode in which the calibration of test specimens is carried out.
(
§ 5.2).
Set point mode:
The set temperatures can be entered in this mode (
§ 5.3).
Main menu:
All the settings can be carried out in this mode, e.g. pre-setting the set temperatures or
setting the control parameters. (
§ 5.4).