SC12205-K10-15002A
2nd
9
Copyright 2020, SIGMAKOKI Co.,Ltd.
Chapter 2: Basic Operations
2-1.HSC-103 Connection procedure
First, connect HSC-103 to the motorized stages.
①
Please confirm the power switch of the HSC-103 is turning off.
②
Connect a standard cable (D15RP-CA/D15D15A-CA) to the connector of the motorized stage.
③
Connect the stage to be controlled as the first axis to the STAGE1 connector of the HSC-103
controller. Also connect the stage controlled as the second axis to the STAGE2 connector. Also
connect the stage controlled as the third axis to the STAGE3 connector.
2-2. Connecting to PC and peripheral device
Connect HSC-103 to PC and peripherals (JS-300/301 / JB-400/401 / JD-100/101).
USB interface is used for the connection between the PC and HSC-103.
The USB interface communication parameter at the time of default Value is described below. Please set
the configurations of the PC side according to the following table.
Parameter
Descriptions
Baud rate
38400[bps]
Delimiters
CR+LF
Parity
None
Data bits
8bit
Stop bit
1bit
Flow control
Hardware (RTS/CTS) or none (Default value is Hardware)
①
Please confirm the power switch of the HSC-103 is turning off.
②
Use a genuine USB cable,
③
Insert of USB cable to the USB connector on the HSC-103.
Please connect peripheral device according to the procedure from
④
to
⑥
if you use it.
④
Please use a special cable MDR14-CA for the connection of peripheral device.
⑤
The one side of MDR14-CA is connected with the connector of the peripheral device.
⑥
The connector on the other side of MDR14-CA is connected with the OPTION connector of
HSC-103.