6 Using the centrifuge
52
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
Translation of the original operating manual
Restrictions
• Deceleration and acceleration curves may include intervals with a
positive slope as well as with a zero slope.
• The slope of the curve intervals may be between 1 rpm/sec and
1000 rpm/sec maximum.
• Square curve intervals are only possible between 0 and 1000 rpm max.
If the selected final speed is higher than 1000 rpm, this interval will
automatically turn into a linear one above 1000 rpm.
• The possible runtime depends on the maximum possible speed
(depending on the rotor) and the slope limitation.
Example 1: Start speed 0 rpm, final speed 100 rpm, a runtime of 1 hour
is not possible since the required slope is < 0.03 rpm and, therefore, is
outside of the defined range.
Example 2: Start speed 0 rpm, final speed 15000 rpm, a runtime of 10
seconds is not possible since the required slope is 1500 rpm and,
therefore, is outside of the defined range.
Pos: 158 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 159 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0020 -0 090 Me nü Hilf e @ 1 0\m od_ 132 8697 107 366 _68 .docx @ 58 100 @ 4 @ 1
6.3.2.9 Help menu
The help menu provides short descriptions of the control elements of the
selected option.
Activating/deactivating the help function
• Select the question mark in the menu bar and press the function knob.
• Quit the help function by selecting the question mark and by pressing
the function knob again.
Parameters can be changed when the help function is activated.
Fig. 25: Help menu
Pos: 160 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1