8KS
8KBS
2 Layout and mode of operation
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
15
Translation of the original operating manual
Pos: 28 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 20 Auf bau und Wirku ngsweis e/02 0-0 020 -00 20- 001 0 D rehz ahl, R adius, R elativ e Z entrif ugal beschl euni gun g @ 7 \mo d_1 306 836 6929 81_ 68. docx @ 44 032 @ 4 @ 1
2.2.2.1 Speed, radius, and relative centrifugal force
The acceleration g, which the samples are subject to, can be increased by
increasing the radius in the rotor chamber and by increasing the speed.
These three parameters are interdependent and linked with each other via
the following formula:
Relative centrifugal force RCF = 11.18 x 10
-6
x r x n
2
r = radius in cm
n = speed in rpm
RCF without any dimension
If two values are entered, the third value is determined by way of the stated
formula. If, afterwards, the speed or the radius is changed, the resulting
relative centrifugal force will be recalculated automatically by the control
unit. If the RCF is changed, the speed will be adapted while the specified
radius is maintained.
The speed-gravitational-field-diagram provides an overview of the
relationship between speed, radius, and RCF (see chapter 11.2 - "
Speed-
gravitational-field-diagram
").
Pos: 29 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 30 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 20 Auf bau und Wirku ngsweis e/02 0-0 020 -00 20- 002 0 Dic hte @ 7\ mod_ 130 683 673 860 6_6 8.docx @ 4 403 9 @ 4 @ 1
2.2.2.2 Density
The laboratory centrifuge is suitable for the separation of constituents of
different densities in mixtures with a maximum density of 1.2 g/cm
3
. All
information concerning the speed of rotors and accessories refers to liquids
with a density corresponding to this specification. If the density is above this
value, the maximum permissible speed of the centrifuge must be reduced
based on the following formula:
n = n
max
x
)
/
2
,
1
(
Rho
Rho = density in g/cm
3