![Sigma 8KBS Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/sigma/8kbs/8kbs_operating-manual_1267914043.webp)
8KS
8KBS
6 Using the centrifuge
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
43
Translation of the original operating manual
Progr.: program selection list
This field in the standard menu shows the program that is currently loaded.
When the field is selected, the program list is displayed (for information on
how to work with the programs, see chapter 6.3.3 - "
Program mode
").
The program "RAPID_TEMP" cannot be deleted (see below).
Fig. 19: Program selection list
Program "RAPID_TEMP"
Precooling at a standstill may distort the measurement results and
subsequently cause increased wear of the mechanical components. This is
why the centrifuge has a special program that precools the centrifuge
rapidly under defined conditions:
• Select the option “Progr” in the “Standard” menu and confirm the
selection. The program list will be displayed.
• Select the program "RAPID TEMP" on the program list and confirm the
selection. The display shows of the maximum rotor speed and the
corresponding RCF value. The deceleration (brake) and acceleration
curves correspond to curve 9 and the runtime field indicates “infinite”
(continuous run).
Fig. 20: "RAPID_TEMP" program
NOTE
The program will only be loaded if the actual temperature is above the set
temperature.