6 Using the centrifuge
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
37
Translation of the original operating manual
Pos: 136 /10 0 Sigm a/11 0 BA Z entrif uge n (Pr ojekt e)/8KS_ 8KBS/060 Bet rieb /06 0-0 020 -00 20- 0060 Blutb eut elsyste me @ 7\m od_ 130 743 366 441 9_6 8.docx @ 4 435 7 @ 4 @ 1
6.2.2.6 Blood bag systems
• All six places on the rotor must be loaded with buckets.
• It is required to fill just two opposite buckets must be filled with one
adapter for blood bags and two blood bag systems each.
• The opposite buckets, including the filled blood bag systems, must have
an equal weight. If the number of blood bags is uneven, a balance
adapter (part no. 17765 for 13865, part no. 17766 for 13866) must be
used. For taring, several balance weights are available (see chapter
11.1 - "
Range of accessories
").
• The blood bags must be put into opposite buckets in a mirror-inverted
way (see figure).
• In both adapter compartments, the main blood bag must be situated
towards the middle. The opposite bucket must be loaded
correspondingly (see figure, item 1).
• When using smaller bag systems or in the case of incompletely filled
blood bags, adapters (e.g. part no. 17750) must be inserted together
with the blood bag systems. This will help to avoid any slipping of the
bags, which could result in an imbalance.
1 Proper loading
2 Possible loading
3 Improper loading
Fig. 14: Loading of blood bag systems
WARNING
The service life of the adapters for blood bags no. 13864, 13865 and
13866 is limited. Refer to the table of rotors and accessories with a
different service life (see chapter 11.4 - "
Table of the service life of
rotors
and accessories ")!