8KS
8KBS
5 Set-up and connection
28
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
Translation of the original operating manual
Pos: 99 / 100 Sigma /110 BA Ze ntrifu gen (Proj ekte )/8KS_8KBS/05 0 Aufst ellun g un d Anschl uss/0 50- 001 1-0 010 Aufstell ung @ 7\ mod _13 074 331 575 49_6 8.d ocx @ 442 35 @ 3 @ 1
5.2.1 Set-up
• Transport the centrifuge as closely as possible to the installation site
with a lifting device (e.g. forklift).
• Set the centrifuge down.
• To place the centrifuge on the castors, open the front door with the
supplied square spanner (part no. 930 114) by turning it clockwise by
90°. Open the front door to the left in order to access the two locking
screws located at the front.
• Loosen the two hexagon lock nuts with the open spanner, size 24 (part
no. 930 024), and screw in the locking screws with the open spanner,
size 8/10 (part no. 930015), in the anticlockwise direction up to the stop
until the adjustment feet are completely relieved of the load.
• Loosen the lock nuts of the two locking screws on the back from the
side and from below with the open spanner, size 24 (part no. 930 024).
Then, screw in the locking screws anticlockwise with the open spanner,
size 13/14 (part no. 930 013), until the adjustment feet are completely
relieved of the load.
• Transport the centrifuge on the castors to the installation site.
CAUTION
Do not use the lid handle to move the centrifuge because it could break
off!
CAUTION
The castors of the centrifuge are made of steel without any plastic coating.
Damage to the surface of the floor cannot be excluded.
Pos: 100 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 101 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 102 /10 0 Sigm a/11 0 BA Z entrif uge n (Pr ojekt e)/8KS_ 8KBS/050 Auf stellu ng u nd Ansc hluss/ 050 -00 11- 0020 Aus richtu ng @ 13\ mod _13 414 708 168 59_6 8.d ocx @ 684 31 @ 3 @ 1
5.2.2 Alignment
DANGER
For reasons of safety, the centrifuge must not be operated while it is set up
on its castors!
NOTE
The centrifuge must be set-up stably and horizontally at the installation
site.
• Set the centrifuge on its adjustable feet (see (see chapter 5.2.1 - "
Set-
up
") in reverse order) To do so, turn all of the feet clockwise by hand
until they touch the ground. Then, perform approximately two more
clockwise turns with the open spanner until the castors are suspended
in the air.