PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
91
pwrup_v
Power-Up Voltage
Parameter
Definition
Acceptance Range
Vport
PSE output voltage following power-up
44 to 57 V
Tested
Parameters
Vpp(low)
Peak to Peak ripple voltage (frequency band 16 Hz –
5K Hz)
0 to 500 mVpp
Optional Input
Controls
Shutdown
Description
Test powers up the PSE and measures the average DC and peak-to-peak AC (low frequency)
voltage after power-up at 2 power extremes. DC Voltage Vport is averaged over the 2 power
extremes and worst case AC Pk-Pk is selected from both power levels. May be used to assess 60
Hz ripple or impact of AC MPS signal on power rail. Low load used is 0 mA for AC MPS PSE’s and
12 mA for DC MPS. High load is 275 mA. User may request to have the test remove power upon
completion of the test.
Trace Analysis
None
Approximate Test
Time
12
seconds per port
Application Note
pwrup_noise
Power-Up High-Frequency Noise
Parameter
Definition
Acceptance Range
Tested
Parameters
Vpp(high)
Peak to Peak ac voltage (frequency band 5Hz –
300K Hz)
0 to 200 mVpp
Optional Input
Controls
Shutdown
Description
Test powers up the PSE and measures the peak-to-peak AC (high frequency) voltage after power-up
at 2 power extremes. Worst case AC Pk-Pk is selected from both power levels. Test may be used
to assess impact of switcher power supply noise. Low load used is 0 mA for AC MPS PSE’s and 12
mA for DC MPS. High load is 275 mA. User may request to have the test remove power upon
completion of the test.
Trace Analysis
None
Approximate Test
Time
5
seconds per port
Application Note
pwrup_pwr
Power-Up Power Level and Current Level
Parameter
Definition
Acceptance Range
Pport
PSE output power following power-up
2.2 to 2.9 Watts
Tested
Parameters
Iport
PSE output current following power-up
49 to 51 mA
Optional Input
Controls
Current_Load | Power, Shutdown
Description
Test powers up the PSE and measures average power and port current. Default condition with no
user input is 50 mA of port current following a Class 0 power-up. User may specify EITHER the
current load following power-up or the target port power. This test can be primarily used as a means
to assure a port power-up to a specified power level. The acceptance criteria are Pass/Info since no
802.3af conformance aspects are strictly tested. User may request to have the test remove power
upon completion of the test.
Trace Analysis
None
Approximate Test
Time
6
seconds per port
Application Note
Содержание PowerSync PSA100
Страница 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Страница 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Страница 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Страница 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Страница 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...