PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
77
4.13. PowerShell PSE Conformance Test Commands
Note
:
The following commands are only available when connected to PSA chassis’ that are enabled for PSE
Conformance Test Suite.
The PSE Conformance Test Suite is addressed in Chapter 5 and was introduced above in Section 3.2. Each PSE
Conformance Test is readily executable from PowerShell utilizing exactly the same test name as it appears in the PSE
(Conformance) Tests menu as well as in the PSE Conformance Sequencer menu.
All PSE Conformance Tests accept the following (optional) command arguments:
-?
Help on command usage and arguments
-v
Verbose diagnostic output as the test executes
-csv
Output to PowerShell will be formatted for copy-paste to a spreadsheet
PowerShell PSE Conformance Test commands are:
Command
Command Parameters
<
port
>
<
r
resistance
> <
c
capacitance
> <
a
aperture
> <
valid
>
det_v
Runs
det_v
test. Optional input Arguments:
r
Specify non-default valid detection signature resistance
resistance
Configure Detection Resistance for Vvalid
RANGE: 9 to 39 DEFAULT: 23 UNIT: K
Ω
c
Specify non-default valid detection signature capacitance
capacitance
Configure Detection Capacitance for Vvalid.
RANGE: 0, 5, 7, or 11 DEFAULT: 0 UNIT:
μ
F
a
Specify fixed measurement timing aperture
aperture
Detection pulse aperture to use.
RANGE: 50m, 100m, 200m, 500m, 1s. DEFAULT: Auto
valid
Return just Vvalid(Max) and Vvalid(Min) values.
<
port
>
det_i
Runs det_i test.
<
port
>
<
bkoff
> <
tdet
> <
det
> <
det_R
>
det_time
Runs
det_time
test. Optional input Arguments:
bkoff
Return only detection backoff parameters
tdet
Return only the 802.3af Tdet Measurement.
detpulse
Return only the total detection pulse time.
det_R
Detection Signature (Kohm) for
tdet
and
detpulse
measurements (Default = ‘25’ K
Ω
)
<
port
>
<
max_time
>
det_range
Runs
det_range
test. Optional input Arguments:
max_time
Maximum Time to Wait for a Power-Up.
Default is ~15 seconds. Range 2 to 30 seconds.
<
port
>
det_rsource
Runs det_rsrouce test.
<
port
>
<
i
current
>
class_v
Runs
class_v
test. Optional input Arguments:
i
Specify non-default class current
current
Classification current load. RANGE: 0 to 45 mA UNITS: mA DEFAULT: 2 mA.
<
port
>
<
i
current
>
class_time
Runs
class_time
test. Optional input Arguments:
i
Specify non-default class current
current
Classification current load. RANGE: 0 to 45 mA UNITS: mA DEFAULT: 2 mA.
Содержание PowerSync PSA100
Страница 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Страница 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Страница 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Страница 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Страница 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...