51
Possible Cause of Error Code
1.
The pressure calibration is incorrect/ corrupted.
Solution:
Contact Sierra for remote troubleshooting.
If this is not immediately possible, pressure may be simulated in the Fluid
Properties sub-menu.
2.
The sensor has failed.
Solution:
Check the Low/High sub-menu (see Table 8) to see if the meter’s pressure limits
have been exceeded.
The meter may need to be
. If this is not immediately
possible, pressure may be simulated in the simulate submenu.
3.
Remote wiring is incorrect
Solution:
Check that remote wiring is as shown
in Figure 19.
Pressure XDCR Overrange
Since mass flow is determined by calculating density from Temperatur
e
and
Pressure
this may
cause errors in the flow readings as well.
240i/241i VTP meters have one transducer for measuring process pressure.
Pressure Ranges:
MP0: No pressure input (incompressible liquids)
MP1: 30 psia
MP2: 100 psia
MP3: 300 psia
MP4: 500 psia
Possible Cause of Error Code
1.
Process pressure exceeded pressure sensor limits.
Solution:
Adjust process conditions to remain within pressure sensor range.
Remove meter from process until pressure is within sensor range.
2.
The pressure calibration is incorrect/ corrupted.
Solution:
Verify wiring, especially if unit is remote. See Figure 19.
Contact Sierra for remote troubleshooting.
If this is not immediately possible, pressure may be simulated in the Fluid
Properties sub-menu.
Содержание InnovaMass 240i Series
Страница 85: ...85 Appendix A Product Specifications ...
Страница 86: ...86 ...
Страница 87: ...87 ...
Страница 88: ...88 ...
Страница 89: ...89 ...
Страница 90: ...90 ...
Страница 91: ...91 ...
Страница 92: ...92 ...
Страница 93: ...93 ...
Страница 94: ...94 ...
Страница 95: ...95 ...
Страница 96: ...96 ...
Страница 111: ...111 ...
Страница 112: ...112 ...
Страница 113: ...113 ...