105
4.2.
Mounting, Commissioning, and Operation
The device has been designed to operate safely in accordance with the current technical and
safety regulations of the EU. If installed incorrectly or used for applications for which it is not
intended, it is possible that application related changes may arise. For this reason, the instrument
must be installed, connected, operated, and maintained according to the instructions in this and
the specific product operating manual.
Persons handling/installing or commissioning this equipment must be authorized and suitably
qualified. The manual must be read, understood, and the instructions must be followed.
Modifications and repairs to the device are only permissible when they are expressly approved in
this manual.
4.3.
Explosive Hazardous Area
If the device is to be installed in an explosive hazardous area, then the specifications in the
certificate as well as all national and local regulations must be observed.
The instrument will be delivered with the certified ATEX/IECEx marking :
ATEX
II 2 G Ex db IIC T3 Gb
II 2 D Ex tb IIIC T200°C Db
IECEx
Ex db IIC T3 Gb
Ex tb IIIC T200°C Db
The certificate type can be identified from the second group of numbers (Feature 2: Approvals)
on the model code stamped on the nameplate. For example:
1= NAA : Not suitable for hazardous areas, Non-Agency Approved.
2= cFMus : Explosion proof for Class I, Division 1, Groups B, C, D
3= ATEX/IECEx:
ATEX
II 2 G Ex db IIC T3 Gb
II 2 D Ex tb IIIC T200°C Db
IECEx
Ex db IIC T3 Gb
Ex tb IIIC T200°C Db
This manual addition only applies to ATEX/IECEx units.
The unit is supplied without cable glands for the power and signal. It is the user’s responsibility to
select suitable cable glands that meet or exceed the required ATEX/IECEx approval and that are
suitable for the signal and power cable used. The connections on the electronics housing for the
input power and signal cable glands is ¾”-14 female NPT threads according to the NPT
requirements of ANSI B1.20.1 plus +0.5 to +2.0 turns deeper.
WARNING: UNUSED WIRING ENTRIES ARE TO BE CLOSED USING SUITABLY
CERTIFIED PLUGS TO MAINTAIN THE ENCLOSURE TYPE OF PROTECTION
Содержание InnovaMass 240i Series
Страница 85: ...85 Appendix A Product Specifications ...
Страница 86: ...86 ...
Страница 87: ...87 ...
Страница 88: ...88 ...
Страница 89: ...89 ...
Страница 90: ...90 ...
Страница 91: ...91 ...
Страница 92: ...92 ...
Страница 93: ...93 ...
Страница 94: ...94 ...
Страница 95: ...95 ...
Страница 96: ...96 ...
Страница 111: ...111 ...
Страница 112: ...112 ...
Страница 113: ...113 ...