43
NMPS, NMPM, NMPH, NMPD, NMPY
Normal Meters per Second, Minute, Day, Hour or Year. Normal conditions
are set in the Reference Conditions menu. Note this is a point VELOCITY.
BTU
British Thermal Units per Second, Minute, Hour, Day, or Year. Reference
conditions N/A.
KJoules
KJoules per Second, Minute, Hour, Day, Year. Reference conditions N/A.
Calibration (Sub-
Menu) “Flow Units”
Available Choices
Volumetric Flow Units
Description
ACFS, ACFM, ACFH, ACFD, ACFY
Actual Cubic Feet per Second, Minute, Day, Hour or Year. No Reference
Conditions are used.
AM3/sec, AM3/min, AM3/hr, AM3/day,
AM3/yr
Actual Cubic Meters per Second, Minute, Day, Hour or Year. No
Reference Conditions are used.
ALPS, ALPM, ALPH, ALPD, ALPY
Actual Liters per unit time. No Reference Conditions are used.
Gal/sec, Gal/min, Gal/hr, Gal/day.
Gal/yr
US Gallons per unit time. No Reference Conditions are used.
MilG/sec, MilG/min, MilG/hr, MilG/day,
MilG/yr
Million US Gallons per unit time. No Reference Conditions are used.
ImpG/sec, ImpG/min, ImpG/hr,
ImpG/day, Imp /yr
Imperial Gallons per unit time. No Reference Conditions are used.
bbl/sec, bbl/min, bbl/hr, bbl/day, bbl
/yr
Barrels per unit time. No Reference Conditions are used.
lit/sec, lit/min, lit/hr, lit/day, lit/yr
Liters per unit time. No Reference Conditions are used. Same as
ALP*.
MilL/sec, MilL/min, MilL/hr, MilL/day,
MilL/yr
Million Liters per unit time. No Reference Conditions are used.
m3/sec, m3/min, m3/hr, m3/day, m3/yr
Meter cubed per unit time. No Reference Conditions are used. Same as
Actual Cubic meters (
AM3/*).
ft3/sec, ft3/min, ft3/hr, ft3/day, ft3/yr
Feet cubed per Second, Minute, Day, Hour or Year. No Reference
Conditions are used. Same as Actual Cubic Feet, (
ACF*)
FPS, FPM, FPH, FPD, FPY
Feet per Second, Minute, Day, Hour or Year. No Reference Conditions
are used. Note this is a point VELOCITY.
MPS, MPM, MPH, MPD, MPY
Meters per Second, Minute, Day, Hour or Year. No Reference Conditions
are used. Note this is a point VELOCITY.
bl/sec, bl/min, bl/hr, bl/day, bl/yr
Beer barrels per Second, Minute, Day, Hour, or Year.
In/sec, In/min, In/hr, In/day, In/yr
Inches per Second, Minute, Hour, Day, or Year.
You can press the exit
several times to climb back up to the main menu level or press
to
proceed to the next Calibration Sub-Menu.
Calibration (Sub-Menu): Span Adjust
The Span Adjust multiplies the meter output by a fixed value, from 0.5000000 to 2.000000.
Displayed Span Adjust is for the currently selected gas. This is useful if the meter displays a
known offset from another device or expectation (calculated flow rate for example), and you want
to make the meters “match”.
For instance, assume your process uses a less accurate rotameter that reads 100 scfm, and the
more accurate 240i or 241i that reads 95 scfm to align. Enter a span factor of 100/95 = 1.052632.
The meter will then multiply by this factor so that the meter output equals the rotameter output of
100 scfm. Default is 1.000000.
MeterSpan
1.00000000
Содержание InnovaMass 240i Series
Страница 85: ...85 Appendix A Product Specifications ...
Страница 86: ...86 ...
Страница 87: ...87 ...
Страница 88: ...88 ...
Страница 89: ...89 ...
Страница 90: ...90 ...
Страница 91: ...91 ...
Страница 92: ...92 ...
Страница 93: ...93 ...
Страница 94: ...94 ...
Страница 95: ...95 ...
Страница 96: ...96 ...
Страница 111: ...111 ...
Страница 112: ...112 ...
Страница 113: ...113 ...