2 S a f e t y a n d C a r e
[ 1 ] I N S T R U C T I O N S F O R U S E
2 - 3 3
Transducer Accessories
Care
WARNING:
Ensure the accessories for a transducer are properly cleaned,
sterilized or disinfected as appropriate before each use to avoid possible
patient contamination.
Instructions for care are provided for the following accessories for
transducers:
Transducer Sheaths
Gel pads
Needle Guide Bracket Kits
Transducer Sheaths
WARNING:
There have been reports of severe allergic reactions to medical
devices containing latex (natural rubber). Health care professionals are advised to
identify latex-sensitive patients and to be prepared to treat allergic reactions
promptly. For additional information in the U.S.A., refer to FDA Medical Alert
MDA91-1.
WARNING:
To minimize the risk of cross-contamination and infectious
diseases, endocavity transducers must be cleaned and high-level disinfected
after each use. A sterile, non-pyrogenic transducer sheath must be in place
during procedures requiring sterility.
Caution:
Siemens recommends that you follow all instructions provided by
manufacturers of sterile goods (transducer sheaths) to ensure proper handling,
storage, and cycling of all sterile goods.
Using a disposable latex transducer sheath on a transducer reduces the
possibility of cross-contamination. Always use a protective transducer
sheath for endocavity exams, and when scanning an open wound or an area
where the skin is not intact.
Storage
Caution:
Do not store transducer sheaths in direct sunlight, as ultraviolet
damage can result.
Latex products have a limited shelf life, and should be stored in a cool, dry,
dark place with an ambient temperature between –5°C and +40°C and up to
80% relative humidity at +40°C. Before use, examine these products for any
material flaws. Some packaging may list an expiration date. Any product
showing flaws, or whose expiration date has passed, should not be used.
[2] Instructions for Use
Attachment
Procedures
Ch C1
MPT7-4
Bite Guard
Ch C3
Содержание SONOLINE G50
Страница 9: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 19: ...1 Introduction 1 12 1 I N S T R U C T I O N S F O R U S E ...
Страница 57: ...2 Safety and Care 2 38 1 I N S T R U C T I O N S F O R U S E ...
Страница 143: ...3 System Controls 3 86 1 I N S T R U C T I O N S F O R U S E ...
Страница 145: ...4 System Setup 4 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 183: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 229: ...6 Technical Description 6 28 1 I N S T R U C T I O N S F O R U S E ...