2 S a f e t y a n d C a r e
[ 1 ] I N S T R U C T I O N S F O R U S E
2 - 1 1
Transmit Power Display
The transmit power range is from –20 dB to 0 dB. Selecting 0 dB, or 100%, in
combination with other system controls or functions generates the maximum
acoustic intensity and mechanical index for each transducer, where:
I
SPTA.3
:
d
720 mW/cm
2
and
MI
d
1.9
Use the system presets to determine whether the power value displays on
the image screen in decibels,
dB
, or as a
percentage
of the maximum
allowable power.
0 dB
-3 dB
-6 dB
-9 dB
-12 dB
-15 dB
-18 dB
-21 dB
-24 dB
100%
71%
50%
35%
25%
18%
13%
9%
6%
Comparison of decibels to percentages.
When the power level is changed, the system briefly highlights the power
value to indicate the change.
C 5/2/ 26
+
+
+
+
.... .... ....
+
+
+
+
.... .... ....
.
.
.
.
.
.
.
.
.
.
.
.
....
....
Location of power display on the image screen.
System Reference
System Presets
Ch 3
F4
General
Ź
Beep Volume
Ź
Tx Power
Display Format