
5 B e g i n n i n g a n E x a m i n a t i o n
5 - 1 8
[ 1 ] I N S T R U C T I O N S F O R U S E
Selecting an Operating Mode
When powered on, the system automatically displays in 2D-mode. You can
change modes by pressing the appropriate control or key located on the
control panel. Selection of A-mode, full-screen M sweeps, and full-screen
spectral Doppler waveforms is accomplished using LCD selections.
Use the system presets to display the 2D image simultaneously with the
M-mode sweep or Doppler spectrum when M-mode or Doppler is selected.
If you do not activate the bypass cursor display selection in the system
presets, the system displays the 2D image with the cursor when M-mode or
Doppler is selected and you must select the control a second time to display
the sweep or spectrum.
Mode
Selection
A-mode
Press the
2D
control on the control panel, push the
PAGE
control on the LCD panel
to display LCD page 2, and then press the
|
A-mode
LCD selection.
2D-mode
Press the
2D
control on the control panel.
Split (B+B) mode
Press the
SPLIT
key on the control panel.
Dual-mode
Press the left or right
DUAL/SELECT
key on the control panel. A 2D image displays in
the respective half of the image area. Press the other
DUAL/SELECT
key to display
an image in the other half of the display area. The last
DUAL/SELECT
key pressed
(left or right) establishes the active image. Press either key twice to display the image
full-screen. Return to dual displays by pressing the key once again.
4B-mode
(Requires software version 1.2 or higher)
Press the
4B
key on the control panel.
2D/M-mode
Press the
M
control on the control panel two times while in 2D-mode.
Split 2D/M-mode
Press the
SPLIT
key on the control panel and then press the
M
control.
M-mode
Press the
M
control on the control panel two times and then press the
|
Full M
LCD
selection to
On
to display full-screen M-mode.
Doppler
Press the
D
control on the control panel two times.
To display full-screen Doppler, press the
|
Full D
LCD selection.
An update function is available to toggle a frozen 2D image with a real-time spectrum
or a real-time 2D image with a frozen spectrum. During 2D-mode with M/D-mode,
the M-mode sweep is synchronized with the Doppler spectrum.
Steerable Continuous
Wave Doppler
1
Connect a phased array transducer to an array port and then press the
CW
key.
Auxiliary Continuous
Wave Doppler
2
Connect a continuous wave (pencil) transducer to the CW port and then press the
CW
key.
Color
During 2D-mode, 2D/M-mode, or 2D with Doppler, press the
C
control.
Power
During 2D-mode, 2D with Doppler, or 2D with Color, press the
POWER
key.
Directional Power
Doppler
Use the
|
Dir Power
Power LCD selection to aid in determining power flow relative
to the transducer.
Press
|
Dir Power
until the required setting displays on the LCD.
1
Available with the SONOLINE G60 S system; requires the Cardiac Imaging option for the
SONOLINE G50 system
2
Available for the SONOLINE G60 S system only
F4
Default Settings
Ź
Bypass M/D
Cursor Display
Содержание SONOLINE G50
Страница 9: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 19: ...1 Introduction 1 12 1 I N S T R U C T I O N S F O R U S E ...
Страница 57: ...2 Safety and Care 2 38 1 I N S T R U C T I O N S F O R U S E ...
Страница 143: ...3 System Controls 3 86 1 I N S T R U C T I O N S F O R U S E ...
Страница 145: ...4 System Setup 4 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 183: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Страница 229: ...6 Technical Description 6 28 1 I N S T R U C T I O N S F O R U S E ...