Operation requiring verification
12.2 Verification
SIWAREX WP231
160
Manual, 07/2014, A5E31238908A-02
12.2.2
Checking of parameters relevant to the device
The following parameters relevant to the device are checked during the verification:
1.
Checking of the software ID of the SecureDisplay
The software ID of the DisplayCali.exe function must agree with the requirements in the
SIWAREX SecureDisplay test certificate (EC). The currently valid version can be called
on the SecureDisplay.
–
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
–
Select the "Version and variables check" submenu in the "Calibration check"
submenu.
–
Activate the "Display SecureDisplay version" command.
–
The version of the SecureDisplay is then shown on the display.
2.
Checking of the firmware ID of the SIWAREX WP231 evaluation electronics
The firmware ID of the SIWAREX WP231 evaluation electronics must agree with the
requirements in the EC construction license. The currently valid version can be called on
the SecureDisplay.
–
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
–
Select the "Version and variables check" submenu in the "Calibration check"
submenu.
–
Activate the "Display SIWAREX firmware release" command.
–
The current SIWAREX firmware release is then shown on the display.
3.
Checking of the smallest zoom factor for the SecureDisplay display software
The zoom factor entered in DR3 for the alternative display size must comply with the
minimum readability and font requirements of EN 45501, Chapter 4.2.1.
The main display with the minimum zoom size can be checked for readability.
–
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
–
Select the "Version and variables check" submenu in the "Calibration check"
submenu.
–
Activate the "Display smallest SecureDisplay" command.
–
The smallest main display is then output and can be checked for readability.
Содержание SIWAREX WP231
Страница 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Страница 24: ...Application planning 4 2 Parameter assignment options SIWAREX WP231 24 Manual 07 2014 A5E31238908A 02 ...
Страница 108: ...Scale parameters and functions 8 15 DR 15 tare settings SIWAREX WP231 108 Manual 07 2014 A5E31238908A 02 ...
Страница 131: ...Scale parameters and functions 8 26 DR 48 date and time 2 for Modbus SIWAREX WP231 Manual 07 2014 A5E31238908A 02 131 ...
Страница 132: ......
Страница 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Страница 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Страница 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...