Commissioning
7.6 5 minute quick setup with SIWATOOL
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
53
7.6.3
Input of required parameters
For commissioning, you must enter the following parameters in data record DR 3 and send
these to the module:
●
Unit of weight
●
Required maximum weighing range of the scale
●
Scale interval
Sending/receiving a data record is always carried out by right-clicking on the data record
name in the "Value" column in the tree structure.
For example, if data record 3 is to be sent, right-click on "Calibration parameter (DR3)". A
submenu is then opened with the option for sending the respective data record to the
weighing module or for reading it from the module. All data records can only be sent as
complete packets to the SIWAREX or read from it. It is not possible to read or write individual
parameters within a data record. Therefore the complete data record must initially be
received for every change to parameters within it. The desired parameter can then be edited,
and the data record returned. If the data record is not received, the danger exists that
different offline parameters will be sent to the scale and overwrite previously active and
intentionally defined parameters.
Figure 7-9
Sending/receiving a data record with SIWATOOL V7
You must subsequently enter the required parameters in data record DR 10 and send these
to the module.
●
Number of support points
●
Characteristic value of a load cell in mV/V, or the mean value of the characteristic values
if there is more than one load cell
●
Rated load of a load cell
7.6.4
Complete automatic calibration
●
The scale must be empty (only mechanical dead load).
●
Activate the "Automatic Calibration 82" command.
●
Activate the "Shift Characteristics 81" command.
Содержание SIWAREX WP231
Страница 16: ...Safety notes 2 1 General safety instructions SIWAREX WP231 16 Manual 07 2014 A5E31238908A 02 ...
Страница 24: ...Application planning 4 2 Parameter assignment options SIWAREX WP231 24 Manual 07 2014 A5E31238908A 02 ...
Страница 108: ...Scale parameters and functions 8 15 DR 15 tare settings SIWAREX WP231 108 Manual 07 2014 A5E31238908A 02 ...
Страница 131: ...Scale parameters and functions 8 26 DR 48 date and time 2 for Modbus SIWAREX WP231 Manual 07 2014 A5E31238908A 02 131 ...
Страница 132: ......
Страница 174: ...Technical data 13 3 Approvals SIWAREX WP231 174 Manual 07 2014 A5E31238908A 02 ...
Страница 180: ...List of abbreviations SIWAREX WP231 180 Manual 07 2014 A5E31238908A 02 ...
Страница 182: ...Index SIWAREX WP231 182 Manual 07 2014 A5E31238908A 02 ...