FM/CSA
SITRANS TH/TR
SITRANS TF
Display
Name
IS,CL I, Div 1, GP ABCD, T6 ... T4
Ex ia IIC T6 … T4 Ga
AEx ia IIC T6 … T4 Ga
or
Ex ib [ia Ga] IIC T6 … T4 Gb
AEx ib [ia Ga] IIC T6 … T4 Gb
IS, CL I, II, III, Div 1, GP ABC‐
DEFG, T6 ... T4
Ex ia IIC T6 ... T4 Ga
AEx ia IIC T6 ... T4 Ga
or
Ex ib [ia Ga] IIC T6 ... T4 Gb
AEx ib [ia Ga] IIC T6 ... T4 Gb
IS,CL I, Div 1, GP ABCD, T6 ... T4
or
Ex ia IIC T6 … T4 Ga
AEx ia IIC T6 … T4 Ga
Output terminals
SITRANS TH/
SITRANS TF single
chamber housing: 1, 2
SITRANS TR: 3, 4
SITRANS TF dual cham‐
ber housing: (+) and
(-)
U
i
= 30 V DC, I
i
= 120 mA, P
i
=
900 mW, L
i
= 0 μH, C
i
= 1.0 nF
U
i
= 30 V DC, I
i
= 120 mA, P
i
=
900 mW, L
i
= 0 μH, C
i
= 2.2 nF
U
i
= 30 V, I
i
= 120 mA, P
i
=
900 mW, C
i
= 2.2 nF, L
i
= 0 µH
• Permissible ambi‐
ent temperature
-50 °C to +85 °C
(-58 °F to +185 °F) temperature
class T4
-50 °C to +70 °C
(-58 °F to +158 °F) temperature
class T5
-50 °C to +55 °C
(-58 °F to +131 °F) temperature
class T6
-50 … +70 °C (-58 … +158 °F)
Temperature class T4
1)
-50 … +40 °C (-58 … +104 °F)
Temperature class T6
1)
-40 … +85 °C (-40 … +185 °F)
Temperature class T4
-40 … +55 °C (-40 … +131 °F)
Temperature class T6
Output terminals
SITRANS TH/
SITRANS TF single
chamber housing: 1, 2
SITRANS TR: 3, 4
SITRANS TF dual cham‐
ber housing: (+) and
(-)
U
i
= 30 V DC, I
i
= 100 mA, P
i
=
750 mW, L
i
= 0 μH, C
i
= 1.0 nF
U
i
= 30 V DC, I
i
= 100 mA, Pi =
750 mW, L
i
= 0 μH, C
i
= 2.2 nF
U
i
= 30 V, I
i
= 100 mA, P
i
=
750 mW, C
i
= 2.2 nF, Li = 0 µH
• Permissible ambi‐
ent temperature
-50 °C to +85 °C
(-58 °F to +185 °F) temperature
class T4
-50 °C to +75 °C
(-58 °F to +167 °F) temperature
class T5
-50 °C to +60 °C
(-58 °F to +140 °F) temperature
class T6
-50 … +75 °C (-58 … +167 °F)
Temperature class T4
1)
-50 … +50 °C (-58 … +122 °F)
Temperature class T6
1)
-40 … +85 °C (-40 … +185 °F)
Temperature class T4
-40 … +60 °C (-40 … +140 °F)
Temperature class T6
Input terminals
SITRANS TH/TF: 3, 4, 5,
6, 7, 8, 9
SITRANS TR: 5, 6, 7, 8,
9, 10, 11, 12
TH/TR/TF320
U
o
= 7.2 V DC, I
o
= 12.9 mA, P
o
= 23.3 mW, L
o
= 200 mH, C
o
=
13.5 μF
-
TH/TR/TF 420 (input 1 or 2)
U
o
= 7.2 V DC, I
o
= 7.3 mA, P
o
= 13.2 mW, L
o
= 667 mH, C
o
=
13.5 μF
• "Non-explosive/ener‐
gy limited NIFW" type
of protection
Technical data
12.2 SITRANS TH320/TH420/TR320/TR420/TF320/TF420
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
210