Getting started
1
1.1
SITRANS TH100 Commissioning
Condition
You have read the following safety instructions:
• General safety notes (Page 27)
• Basic safety notes: Installing/mounting (Page 61)
• Basic safety notes: Connecting (Page 77)
• Basic safety notes: Commissioning (Page 117)
Read the entire device manual in order to achieve the optimum performance of the device.
Procedure
1. If you want to change the factory settings, configure the device using the USB modem and
the SIPROM T parameter assignment software.
Parameter assignment with USB modem and SIPROM T (Page 137)
2. Mount or install the device depending on the application.
Installing SITRANS TH in a connection head (Page 65)
Installing SITRANS TH on DIN rail and G rail (Page 67)
3. Connect the device.
Connecting the TH100 (Page 79)
4. For devices with intrinsically safe protection type, use an isolating power supply that
corresponds to the requirements of the relevant type of protection.
5. Switch on the power supply.
Switching on the supply voltage (Page 128)
6. Wait for 10 seconds.
After this startup time, the device is operational.
7. Wait 5 minutes to get exact measured values.
See also
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
Operating Instructions, 05/2020, A5E41864807-AB